Patient Monitor (12.1 inch)  - Series : Soulful 10003

 

Feature

  • 12.1-inch Six Standard parameters: ECG, RESP, NIBP, SPO2, 2_TEMP, PR/HR
  • 13 Type Arrhythmic Analysis, Multi_Lead ECG Waveforms Display in Phase, Real time S_T segment analysis, pacemaker detection Drug calculation and titratiotable;
  • Efficient resistance to interference of defibrillator and electrosurgical cautery;
  • SPO2 can testing for 0.1% Weak;
  • RA-LL impedance Respiration;
  • Trend Coexist Display;
  • OxyCRG DyNamic View Display;
  • Bed to Bed view Display;
  • Networking capacity and nurse calling system;

10. Options of Central Monitoring System, IBP, EtCO2, Recorder, 12-Channel ECG, Touch Screen, VGA;

11. Built-in rechargeable battery 4400mAh;

12. 12.1" high resolution color TFT LCD display;

13. Large volume of tabular and graphic trends information storage and easy to recall;

14. Anti-ESU, anti- defibrillator;

15. Capture dynamic waveforms.

Patient Monitor (12.1 inch)

SKU: Series : Soulful 10003
$1,100.00 Regular Price
$990.00Sale Price
Quantity
  • Technical Specification

    ECG

    Lead Mode: 5 Leads (R, L, F, N, C or RA, LA, LL, RL, V)

    Lead selection: I, II, III, avR, avL, avF, V,

    Waveform: 2 ch

    Lead mode: 3 Leads (R, L, F or RA, LA, LL)

    Lead selection: I, II, III,

    Waveform: 1 ch

    Gain: ‘2.5mm/mV, ‘5.0mm/mV, ‘10mm/mV, ‘20mm/mV, auto

     

    Heart Rate and Alarm

    Range

    Adult: 15 ~ 300 bpm

    Neo/Ped: 15 ~ 350 bpm

    Accuracy: ± 1% or ± 1bpm, which great

    Resolution: 1 bpm

    Sensitivity: > 200 (uV P-P)

    Differential Input Impedance: > 5 MΩ

    CMRR

    -Monitor: > 105 dB

    -Operation: > 105 dB

    -Diagnosis: > 85 dB

    Electrode offset potential: ±300mV

    Leakage Current: < 10 uA

    Baseline Recovery: < 3 S after Defi.

    ECG Signal Range: ±8 m V (Vp-p)

    Bandwidth

    -Surgery: 1 ~ 15 Hz

    -Monitor: 0.5 ~ 35 Hz

    -Diagnostic: 0.05 ~ 100 Hz

    Calibration Signal: 1 (mV p-p), Accuracy: ±5%

    ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV

    ARR Detecting

    Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC

    Alarm: Available

    Review: Available

     

    Respiration

    Method: Impedance between R-F (RA-LL)

    Differential Input Impedance: >2.5 MΩ

    Measuring Impedance Range: 0.3~5.0Ω

    Base line Impedance Range: 0 – 2.5 KΩ

    Bandwidth: 0.3 ~ 2.5 Hz

    Resp. Rate

    Measuring and Alarm Range

    -Adult: 0 ~ 120 rpm

    -Neo/Ped: 0 ~ 150 rpm

    Resolution: 1 rpm

    Accuracy: ±2 rpm

    Apnea Alarm: 10 ~ 40 S

     

    NIBP

    Method: Oscillometric

    Mode: Manual, Auto, STAT

    Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90,

    120, 180, 240,480 (Min)

    Measuring Period in STAT Mode: 5 Min

    Pulse Rate Range: 40 ~ 240 bpm

    Alarm Type: SYS, DIA, MEAN

    Measuring and alarm range

    Adult Mode

    -SYS: 40 ~ 270 mmHg

    -DIA: 10 ~ 215 mmHg

    -MEAN: 20 ~ 235 mmHg

    Pediatric Mode

    -SYS: 40 ~ 200 mmHg

    -DIA: 10 ~ 150 mmHg

    -MEAN: 20 ~ 165 mmHg

    Neonatal Mode

    -SYS: 40 ~ 135 mmHg

    -DIA: 10 ~ 100 mmHg

    -MEAN: 20 ~ 110 mmHg

    Resolution Pressure: 1mmHg

    Accuracy Pressure Maximum Mean error: ±5mmHg

    Maximum Standard deviation: ±8mmHg

    Overpressure Protection

    -Adult Mode: 297±3 mmHg

    -Pediatric Mode: 240±3 mmHg

    -Neonatal Mode: 147±3 mmHg

     

    SpO2

    Measuring Range: 0 ~ 100 %

    Alarm Range: 0 ~ 100 %

    Resolution: 1 %

    Accuracy: 70% ~ 100% ±2 %; 0% ~ 69% unspecified

    Actualization interval: about 1 Sec.

    Alarm Delay: 10 Sec.

    Pulse Rate

    Measuring and Alarm Range: 20~300bpm

    Resolution: 1bpm

    Accuracy: ±2bpm

     

    Temperature

    Channel: 2

    Measuring and Alarm Range: 0 ~ 50 °C

    Resolution: 0.1°C

    Accuracy: ±0.1°C

    Actualization interval: about 1 Sec.

    Average Time Constant:<10 Sec.

     

    IBP (optional)

    Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2

    Measuring and alarm range

    -ART: 0 ~ 300 mmHg

    -PA: -6 ~ 120 mmHg

    -CVP/RAP/LAP/ICP: -10 ~ 40 mmHg

    -P1/P2: -10 ~ 300 mmHg

    Press Sensor

    Sensitivity: 5 uV/V/mmHg

    Impedance: 300-3000Ω

    Resolution: 1 mmHg

    Accuracy: ±2% or ±1mmHg, which great

    Actualization interval: about 1 Sec

     

    Standard Packing List

    5-lead ECG cable (including RESP)

    Adult finger SpO2 sensor

    Adult NIBP cuff

    NIBP extension tube

    TEMP sensor

    AC power adaptor

    ECG electrodes

    Operation instruction

     

    Optional Function

    Central Monitoring System, IBP, EtCO2, Recorder, 12-Channel ECG, Touch Screen, VGA

     